Ambulatory polysomnography
Vitaport 4 Ambulatory PSG recorder
Channels 1-10
Type
AC-coupled common reference
EEG/EOG/EMG
| AC Input Range |
5000 µVpp |
| DC Input Range |
> ± 200mV (Ag/AgCl electrodes recommended) |
| Low pass filter |
70 Hz, 2nd order butterworth |
| High pass filter |
0,3 Hz (TC = 0,53S.), 1st order RC |
| Noise |
< 0,5 µV RMS, 0.1 - 70 Hz |
| CMRR |
> 120 dB (1 - 60 Hz) with driven neutral |
| Input impedance |
> 100 MΩ |
| Sample rates |
128 Hz |
| Resolution |
16 bits |
| Time skew |
Synchronous sampling |
Nasal pressure
AC nasal pressure derived
from nasal pressure sensor for:
-
nasal pressure
cannula
-
CPAP mask pressure
| Input range: ± 25mbar |
Sample rate: 32 Hz |
| Time constant: 6 sec |
Low pass: 4 Hz |
Snore from NP
AC snore channel derived
from nasal pressure sensor
| Input range: ± 100% |
Sample rate: 128Hz |
| High pass: 4 Hz |
|
CPAP pressure from NP
DC CPAP pressure channel
derived from nasal pressure sensor
| Input range: ± 25mbar |
Sample rate: 4 Hz |
| Low pass: 0.028 Hz |
|
Flow
Flow channel by
thermo-couple sensor
| Input range: ± 0,5 mV |
Sample rate: 64 Hz |
| Time constant: 3 sec |
Low pass: 15 Hz |
Chest excursions
Respiratory Inductive
Plethysmography or Piezo electrical
Abdomen
excursions
| Input range: ± 2,5 mV |
Sample rate: 64 Hz |
| Time constant: 3 sec |
Low pass: 15 Hz |
Body Position
Built-in sensor for
supine, right, prone, left and up positions
| Input range: 0 - 1 V |
Sample rate: 4 Hz |
| Time constant: DC |
Low pass: 30 Hz |
ECG (1, 2, 3)
ECG/EMG/ExG channels (can
also be used for EMG, snoring, activity sensors)
| Input range: ± 5 mV |
Sample rate: 128 Hz |
| Time constant: 0,3 sec |
Low pass: 100 Hz |
| CMRR: > 100 dB |
|
Snore piezo-electrical
Via ECG/EMG/ExG amplifier
Activity
Via ECG/EMG/ExG amplifier
SpO2
Oxygen saturation via
serial Nonin XPOD adapter
Sample rate: 4 Hz
Pulse Waveform
Pulse waveform via serial
Nonin Xpod adapter
Sample rate: 128 Hz
PR
Pulse rate
Sample rate: 4 Hz
Event marker channel
The "ON" button can be
used as marker in channel.
Sample rate: 4 Hz
Battery
monitoring
Internal continuous
battery monitoring channel
Sample rate: 4 Hz
Resolution A/D conv.
16 bits
Dimensions and weight
| 70 x 122 x 25 mm |
145 gram |
Battery
Internal rechargeable
Lithium Ion polymer
Recording duration
> 24hours on one battery
charge
Connectors
1 connector to the
sensors
1 connector to amplifier
box
2 connectors to serial
interfaces (computer)
1 connector for battery
charger/DC-adapter
Storage
Flash card storage in EDF
format
Quality
system
Certification: conform
Directive 93/42/EEC CE0197
Standards: IEC 60601-1
|